Fast transimpedance preamplifier for a boron-coated multiwire proportional chamber neutron detector
Zhang, Y; Yan, HY; Zhang, J; Sun, ZJ; Cao, P; An, Q; Gong, J; Sun ZJ(孙志嘉)
2016
发表期刊EUROPEAN PHYSICAL JOURNAL PLUS
卷号131期号:10
文章类型Article
摘要A low-noise and fast preamplifier is needed for the new boron-coated multiwire proportional chamber (MWPC) neutron detector with a delay line readout system to improve position resolution. A transimpedance preamplifier with a rise time of 30 ns, a signal-to-noise ratio higher than 40 dB, and an automatic gain control function are designed to meet the aforementioned requirements. On the other hand, we also compare it to commercial preamplifiers. Then, the advantages and disadvantages of commercial and in-house preamplifiers are analyzed. The preamplifier is used and tested on a MWPC neutron detector, and results are presented.
学科领域Physics
DOI10.1140/epjp/i2016-16376-y
关键词[WOS]GEM
收录类别SCI
语种英语
WOS类目Physics, Multidisciplinary
WOS记录号WOS:000386675400001
引用统计
文献类型期刊论文
条目标识符http://ir.ihep.ac.cn/handle/311005/260211
专题东莞分部
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Zhang, Y,Yan, HY,Zhang, J,et al. Fast transimpedance preamplifier for a boron-coated multiwire proportional chamber neutron detector[J]. EUROPEAN PHYSICAL JOURNAL PLUS,2016,131(10).
APA Zhang, Y.,Yan, HY.,Zhang, J.,Sun, ZJ.,Cao, P.,...&孙志嘉.(2016).Fast transimpedance preamplifier for a boron-coated multiwire proportional chamber neutron detector.EUROPEAN PHYSICAL JOURNAL PLUS,131(10).
MLA Zhang, Y,et al."Fast transimpedance preamplifier for a boron-coated multiwire proportional chamber neutron detector".EUROPEAN PHYSICAL JOURNAL PLUS 131.10(2016).
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
20160998.pdf(756KB)期刊论文作者接受稿限制开放CC BY-NC-SA请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Zhang, Y]的文章
[Yan, HY]的文章
[Zhang, J]的文章
百度学术
百度学术中相似的文章
[Zhang, Y]的文章
[Yan, HY]的文章
[Zhang, J]的文章
必应学术
必应学术中相似的文章
[Zhang, Y]的文章
[Yan, HY]的文章
[Zhang, J]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。