IHEP OpenIR  > 多学科研究中心
A furnace to 1200 K for in situ heating x-ray diffraction, small angle x-ray scattering, and x-ray absorption fine structure experiments
Cai Q(蔡泉); Wang W(王维); Mo G(默广); Zhang KH(张坤浩); Cheng WD(程伟东); Xing XQ(邢雪青); Chen ZJ(陈中军); Wu ZH(吴忠华); Cai, Q; Wang, Q; Wang, W; Mo, G; Zhang, KH; Cheng, WD; Xing, XQ; Chen, ZJ; Wu, Z
2008
Source PublicationREVIEW OF SCIENTIFIC INSTRUMENTS
Volume79Issue:12Pages:126101
Corresponding Author[Cai, Quan ; Wang, Wei ; Mo, Guang ; Zhang, Kunhao ; Cheng, Weidong ; Xing, Xueqing ; Chen, Zhongjun ; Wu, Zhonghua] Chinese Acad Sci, Inst High Energy Phys, Beijing Synchrotron Radiat Facil, Beijing 100049, Peoples R China ; [Wang, Qiang] Chinese Acad Sci, Inst Phys, Beijing Natl Lab Condensed Matter Phys, Beijing 100080, Peoples R China ; [Cai, Quan ; Wang, Wei ; Mo, Guang ; Zhang, Kunhao ; Cheng, Weidong ; Xing, Xueqing] Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
SubtypeArticle
AbstractA furnace with a water-cooled outside shell has been assembled to do in situ x-ray diffraction (XRD), small angle x-ray scattering (SAXS), and x-ray absorption fine structure (XAFS) experiments. The details of the furnace are described in this paper. The in situ XRD, SAXS, and XAFS experiments during the heating process demonstrate that the available temperature range of this furnace is from room temperature to 1200 K with a temperature accuracy of +/- 0.1 K. By using this furnace, in situ XRD, SAXS, and XAFS experimental techniques with temperature change can be easily combined together.
Keywordcooling furnaces shells (structures) X-ray apparatus X-ray diffraction X-ray scattering
Subject AreaInstruments & Instrumentation; Physics
DOI10.1063/1.3021473
URL查看原文
Indexed ByADS
Language英语
WOS Research AreaInstruments & Instrumentation ; Physics
WOS SubjectInstruments & Instrumentation ; Physics, Applied
WOS IDWOS:000262224800054
ADS Bibcode2008RScI...79l6101C
ADS URLhttps://ui.adsabs.harvard.edu/abs/2008RScI...79l6101C
ADS CITATIONShttps://ui.adsabs.harvard.edu/abs/2008RScI...79l6101C/citations
Citation statistics
Cited Times:3 [ADS]
Document Type期刊论文
Identifierhttp://ir.ihep.ac.cn/handle/311005/240994
Collection多学科研究中心
实验物理中心
Affiliation中国科学院高能物理研究所
First Author AffilicationInstitute of High Energy
Recommended Citation
GB/T 7714
Cai Q,Wang W,Mo G,et al. A furnace to 1200 K for in situ heating x-ray diffraction, small angle x-ray scattering, and x-ray absorption fine structure experiments[J]. REVIEW OF SCIENTIFIC INSTRUMENTS,2008,79(12):126101.
APA 蔡泉.,王维.,默广.,张坤浩.,程伟东.,...&Wu, Z.(2008).A furnace to 1200 K for in situ heating x-ray diffraction, small angle x-ray scattering, and x-ray absorption fine structure experiments.REVIEW OF SCIENTIFIC INSTRUMENTS,79(12),126101.
MLA 蔡泉,et al."A furnace to 1200 K for in situ heating x-ray diffraction, small angle x-ray scattering, and x-ray absorption fine structure experiments".REVIEW OF SCIENTIFIC INSTRUMENTS 79.12(2008):126101.
Files in This Item:
File Name/Size DocType Version Access License
9986.pdf(349KB)期刊论文作者接受稿限制开放CC BY-NC-SAApplication Full Text
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[蔡泉]'s Articles
[王维]'s Articles
[默广]'s Articles
Baidu academic
Similar articles in Baidu academic
[蔡泉]'s Articles
[王维]'s Articles
[默广]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[蔡泉]'s Articles
[王维]'s Articles
[默广]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.