Investigation of misalignment in analyzer crystal based-CT and its effect
Zhang K(张凯); Zhu PP(朱佩平); Huang WX(黄万霞); Yuan QX(袁清习); Li ER(李恩荣); Liu YJ(刘宜晋); Liu XS(刘小松); Wang ZL(王志立); Wu ZY(吴自玉); Zhang, K; Zhu, PP; Huang, WX; Yuan, QX; Li, ER; Liu, YJ; Liu, XS; Wang, ZL; Wu, ZY
刊名PHYSICS IN MEDICINE AND BIOLOGY
2008
卷号53期号:20页码:5757-5766
学科分类Engineering; Radiology, Nuclear Medicine & Medical Imaging
DOI10.1088/0031-9155/53/20/013
通讯作者[Zhang, Kai ; Zhu, Peiping ; Huang, Wanxia ; Yuan, Qingxi ; Li, Enrong ; Liu, Yijin ; Liu, Xiaosong ; Wang, Zhili ; Wu, Ziyu] Chinese Acad Sci, Inst High Energy Phys, Beijing 100049, Peoples R China ; [Wu, Ziyu] Univ Sci & Technol China, Natl Synchrotron Radiat Lab, Hefei 230026, Peoples R China ; [Wu, Ziyu] Chinese Acad Sci, Theoret Phys Ctr Sci Facil, Beijing 100049, Peoples R China
文章类型Article
英文摘要Analyzer crystal-based computed tomography (ACB-CT) is a novel x-ray phase contrast computed tomography technique. In this paper, in order to extract the refraction angle, we analyzed the effect of misalignment on tomography-sliced images of ACB-CT. Two different methods were considered: the first, proposed by Chapman and Dilmanian, was based on two sets of projective image data taken at both sides of the rocking curve and with a half circle sample rotation and the second, recently proposed by our team, was based on only one set of projective image data taken at one side of the rocking curve but with a full circle sample rotation. Theoretical analysis and experimental results will show that the second method improves the quality of reconstructed CT images, also simplifying the ACB-CT data acquisition procedure.
类目[WOS]Engineering, Biomedical ; Radiology, Nuclear Medicine & Medical Imaging
研究领域[WOS]Engineering ; Radiology, Nuclear Medicine & Medical Imaging
原文出处查看原文
语种英语
WOS记录号WOS:000259736400014
引用统计
被引频次:6[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.ihep.ac.cn/handle/311005/240048
专题中国科学院高能物理研究所_多学科研究中心_期刊论文
中国科学院高能物理研究所_多学科研究中心
作者单位中国科学院高能物理研究所
推荐引用方式
GB/T 7714
Zhang K,Zhu PP,Huang WX,et al. Investigation of misalignment in analyzer crystal based-CT and its effect[J]. PHYSICS IN MEDICINE AND BIOLOGY,2008,53(20):5757-5766.
APA 张凯.,朱佩平.,黄万霞.,袁清习.,李恩荣.,...&Wu, ZY.(2008).Investigation of misalignment in analyzer crystal based-CT and its effect.PHYSICS IN MEDICINE AND BIOLOGY,53(20),5757-5766.
MLA 张凯,et al."Investigation of misalignment in analyzer crystal based-CT and its effect".PHYSICS IN MEDICINE AND BIOLOGY 53.20(2008):5757-5766.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
7550.pdf(915KB)期刊论文作者接受稿开放获取CC BY-NC-SA浏览 请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[张凯]的文章
[朱佩平]的文章
[黄万霞]的文章
百度学术
百度学术中相似的文章
[张凯]的文章
[朱佩平]的文章
[黄万霞]的文章
必应学术
必应学术中相似的文章
[张凯]的文章
[朱佩平]的文章
[黄万霞]的文章
相关权益政策
暂无数据
收藏/分享
文件名: 7550.pdf
格式: Adobe PDF
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。