Test of high time resolution MRPC with different readout modes for the BESIII upgrade
Yang, S; Sun, YJ; Li, C; Heng, YK; Qian, S; Chen, HF; Chen, TX; Dai, HL; Fan, HH; Liu, SB; Liu, SD; Jiang, XS; Shao, M; Tang, ZB; Zhang, H; Zhao, ZG;衡月昆; Qian S(钱森); Dai HL(代洪亮); Liu SD(刘曙东); Jiang XS(江晓山)
2014
发表期刊NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
卷号763页码:190-196
摘要In order to further enhance the particle identification capability of the Beijing Spectrometer (BESIII), it is proposed to upgrade the current end-cap time-of-flight (eTOF) detector with multi-gap resistive plate chamber (MRPC). The prototypes, together with the front end electronics (FEE) and time digitizer (TDIG) module have been tested at the E3 line of Beijing Electron Positron Collider (BEPCII) to study the difference between the single and double-end readout MRPC designs. The time resolutions (sigma) of the single-end readout MRPC are 47/53 ps obtained by 600 MeV/c proton/pion beam, while that of the double-end readout MRPC is 40 Ps (proton beam). The efficiencies of three MRPC modules tested by both proton and pion beam are better than 98%. For the double-end readout MRPC, no incident position dependence is observed. (C) 2014 Elsevier B.V. All rights reserved.
关键词Multi-gap resistive plate chamber Single-end readout Double-end readout BESIII End-cap TOF
学科领域Instruments & Instrumentation; Nuclear Science & Technology; Physics; Spectroscopy
DOI10.1016/j.nima.2014.05.094
收录类别SCI
WOS记录号WOS:000341986200027
引用统计
被引频次:8[WOS]   [WOS记录]     [WOS相关记录]
文献类型期刊论文
条目标识符http://ir.ihep.ac.cn/handle/311005/225220
专题实验物理中心
推荐引用方式
GB/T 7714
Yang, S,Sun, YJ,Li, C,et al. Test of high time resolution MRPC with different readout modes for the BESIII upgrade[J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT,2014,763:190-196.
APA Yang, S.,Sun, YJ.,Li, C.,Heng, YK.,Qian, S.,...&江晓山.(2014).Test of high time resolution MRPC with different readout modes for the BESIII upgrade.NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT,763,190-196.
MLA Yang, S,et al."Test of high time resolution MRPC with different readout modes for the BESIII upgrade".NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT 763(2014):190-196.
条目包含的文件
文件名称/大小 文献类型 版本类型 开放类型 使用许可
11243.pdf(3916KB)期刊论文作者接受稿限制开放CC BY-NC-SA请求全文
个性服务
推荐该条目
保存到收藏夹
查看访问统计
导出为Endnote文件
谷歌学术
谷歌学术中相似的文章
[Yang, S]的文章
[Sun, YJ]的文章
[Li, C]的文章
百度学术
百度学术中相似的文章
[Yang, S]的文章
[Sun, YJ]的文章
[Li, C]的文章
必应学术
必应学术中相似的文章
[Yang, S]的文章
[Sun, YJ]的文章
[Li, C]的文章
相关权益政策
暂无数据
收藏/分享
所有评论 (0)
暂无评论
 

除非特别说明,本系统中所有内容都受版权保护,并保留所有权利。