AbstractThe spatial resolution and the relative density resolution are the two most critical indicators in CT system. The method recommended in the ASTM E1695-95 and GJB 5311-2004 is only suitable to the fan-beam CT system. In this paper, for industrial cone-beam micro CT system, we will adopt the edge response function (ERF) created by the step edges of a steel ball to measure the system 3D PSF and MTF. To describe the contrast discrimination function more accurately, we will first propose to extend the two-dimensional measurement region to the three-dimensional space. Our experimental spatial resolution is (55.56+/-0.56) lp/mm and the relative density resolution is 1% within 300 mu mx300 mu mx300 mu m according to the 3 sigma rule.
Keywordmicro computed tomography point spread function contrast discrimination function modulation transfer function
Subject AreaPhysics
DOI10.1088/1074-1137/37/7/078202
Indexed BySCI
Language英语
WOS IDWOS:000322937400017
Citation statistics
Cited Times:2[WOS]   [WOS Record]     [Related Records in WOS]
Document Type期刊论文
Identifierhttp://ir.ihep.ac.cn/handle/311005/224759
Collection核技术应用研究中心
Recommended Citation
GB/T 7714
Wang YF,Que JM,Cao DQ,et al. Measurement of the spatial resolution and the relative density resolution in an industrial cone-beam micro computed tomography system[J]. CHINESE PHYSICS C,2013,37(7):78202.
APA 王燕芳.,阙介民.,曹大泉.,Wang, YF.,Que, JM.,...&魏龙.(2013).Measurement of the spatial resolution and the relative density resolution in an industrial cone-beam micro computed tomography system.CHINESE PHYSICS C,37(7),78202.
MLA 王燕芳,et al."Measurement of the spatial resolution and the relative density resolution in an industrial cone-beam micro computed tomography system".CHINESE PHYSICS C 37.7(2013):78202.
Files in This Item:
File Name/Size DocType Version Access License
10574.pdf(2943KB)期刊论文作者接受稿限制开放CC BY-NC-SAApplication Full Text
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[王燕芳]'s Articles
[阙介民]'s Articles
[曹大泉]'s Articles
Baidu academic
Similar articles in Baidu academic
[王燕芳]'s Articles
[阙介民]'s Articles
[曹大泉]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[王燕芳]'s Articles
[阙介民]'s Articles
[曹大泉]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.